DIGITAL AUTO-TRAK™ SENSITIVITY – Like many other Respironics bi-level ventilators, the BiPAP Focus System includes Respironics’ proprietary Digital Auto-Trak Sensitivity. This feature assures optimum triggering and cycling sensitivity throughout changing breathing patterns and leaks. It also eliminates the need for a perfect seal of the patient interface and constant adjustment while promoting patient-ventilator synchrony.
LARGE USER INTERFACE – Designed for the hospital environment, the user interface is easy to use. It allows the practitioner to easily and quickly adjust patient care settings.
PATIENT PARAMETERS DISPLAY – This valuable feature conveniently allows the practitioner to view all patient parameters at once.
PRESSURE BAR GRAPH – Quickly and clearly indicates patient breaths and pressure ranges.
EASY-TO-USE ALARMS – Pressure alarms are automatically set to 5 cm H2O above IPAP and below EPAP settings, minimizing nuisance alarms and enhancing patient safety.
BACKUP BATTERY – An integral backup battery system maintains patient ventilation in the event of an interruption in power. This safety feature also provides a convenient method for intra-hospital patient transport from one care area to another with uninterrupted ventilatory care.
TOTAL VENTILATION SOLUTIONS – The BiPAP Focus Ventilator is part of our Total Ventilation SolutionsSM that help treat, monitor and manage respiratory-impaired< patients.
Specifications
CPAP Pressure: 4 to 20 cm H2O (4 to 20 hPa) Resolution: 1 cm H2O (1 hPa) Dynamic Accuracy: ±5 cm H2O (±5 hPa)
EPAP Pressure: 4 to 20 cm H2O (4 to 30 hPa) Resolution: 1 cm H2O (1 hPa) Dynamic Accuracy: ±5 cm H2O (±5 hPa)
IPAP Pressure: 4 to 30 cm H2O (4 to 20 hPa) Resolution: 1 cm H2O (1 hPa) Dynamic Accuracy: ±5 cm H2O (±5 hPa)
Rise Time: Settings 1 to 6 (where 1 = 0.1 sec and 6 = 0.6 sec) Resolution: 1 Accuracy: ±10% of setting (at R=50, C=20)
Rate: 1 to 30 BPM
Resolution: 1 BPM Accuracy: ±1BPM or ±10 % of setting whichever is greater over a 4 minute period. I-Time and Rate settings cannot allow I-Time to exceed expiratory time.
I-Time: 0.5 to 3 sec Resolution: 0.1 sec Accuracy: ±0.1+10% of setting sec I-Time and Rate settings cannot allow I-Time to exceed expiratory time.
Ramp Time: 0 to 45 min Resolution: 5 min Accuracy: ±10% of setting
Ramp Start: 4 cm H2O to EPAP (4 hPa to EPAP) Resolution: 1 cm H2O (1 hPa)
Patient Circuit Pressure: Range: 0 to 35 cm H2O (0 to 35 hPa) (continuous bar graph display)
Rate: Range: 0 to 99 BPM
Estimated Vt: Range: 0 to 4000 mL
Estimated MV: Range: 0 to 99 L/min
Patient Leak: Range: 0 to 200 L/min
Apnea Rate: Range: 0 to 99 per hour
High System Pressure Alarms: 5 cm H2O (5 hPa) over target IPAP pressure (Automatic)
Low System Pressure Alarms: 5 cm H2O (5 hPa) under target EPAP pressure (Automatic)
Apnea Alarms: Range: 20, 40, 60 sec or Off
Apnea Rate Alarms: Range: 5, 10, 20 apneas per hour or Off
Nurse Call Alarms: Supports normally open or normally closed systems
Dimensions: 5.5″ H x 11.4″ W x 14″ D (140 mm x 290 mm x 355 mm)
Weight: 10 lb (4.5 kg)
Input Rating: 100VAC / 230VAC and 50/60 Hz
Battery Backup: 3.8 amp-hour (NiMH)
Fuse: No replaceable fuse
Nurse Call: Rated current: 0.280 A Rated voltage: 250 V