Description
Product Specifications
Dimensions
12.0″ x 14.75″ x 7.5″ (30.5 cm x 37.5 cm x 19.1 cm)
Weight
17.9 lb (8.1 K)g
Battery
12 volt, 2.0 ampere, sealed lead acid
Battery Recharge
80% charged within 8 hr.
Battery Life
6 hr at 1-1,400 mL/hr (4 hr with both channels operating at 1-1,400 mL/hr)
Power Cord
Biomed replaceable 9 ft (2.75 m)
Leakage Current
Less than 50 micro-amps (per UL 544)
Fuse
.8 ampere, 250 volt, slow-blow
Communications Port
RS232 serial port
Operating Environment
15.5-37.7ºC (60-100ºF)
Power Requirements
115 VAC, 60 HZ, 50 W
Flow Rate Range
Primary program: 1.0-99.9 mL/hr in 0.1 mL increments or 1-1999 mL/hr in 1 mL increments.Upper limit can be reduced by authorized service personnel.
Secondary program: 1.0-99.9 mL/hr in 0.1 mL increments or 1-999 mL/hr in 1 mL increments
VTBI Range
1.0-99.9 mL in 0.1 mL increments or 1-9999 mL in 1 mL increments for both primary and secondary. Upper limit can be reduced by authorized service personnel.
Ramp PDP Parameter Ranges
Total VTBI: 100-9,999 in 1 mL increments. Upper limit can be reduced by authorized service personnel.
Delay Time: 00:00 – 08:00 hours in minute increments.
Run Time: 01:00 – 16:00 hours in minute increments.
Ramp Up: 00:00 to 45% of the Run Time, in minute increments.
Ramp Down: 00-00 to the Run Time minus Ramp Up, in minute Increments.
Air-in-Line Detection
Factory set to NORM, which causes the pump to alarm on air bubbles approximately 75 uL or larger. The MIN setting causes the pump to alarm on air bubbles approximately 50 ul or larger.
Nominal Downstream
<21
2 PSIG (Minimum)
5 PSIG (Moderate)
9 PSIG (Maximum)
21-200
5 PSIG (Minimum)
8 PSIG (Moderate)
11 PSIG (Maximum)
>200
9 PSIG (Minimum)
12 PSIG (Moderate)
15 PSIG (Maximum)