Description
Standards:
UL 544CSA C22.0, No 125
Case Material:
Impact resistant polycarbonate/ABS alloy
Dimensions:
Height: 7.875 inches (20.00 centimeters)
Width: 6 inches 815.24 centimeters)
Depth: 2.10 inches (5.33 centimeters)
Weight: Approximately 5.1 lbs. (2.32 kg) includes Pole Clamp
Air-In-Line (Default):
500 µl except for Neonatal which is 50µl
Occlusion Pressure (Default):
15 psi except for Controller pressure device which is 3 ft H20
Operating Temperature:
50- 104º Fahrenheit (10º – 40º Celsius)
Storage Temperature:
<95º Fahrenheit (<35º Celsius) for optimum battery life
Maximum Storage Temperature:
131º Fahrenheit (55º Celsius)
Rate Range:
0.1 – 999 milliliter per hour (each channel)
Volume Range:
0.1 – 999 milliliter (each channel)
KVO Rate Range:
0.1 – 20.0 milliliter per hour
System Accuracy:
±5%
Power Consumption:
6 watts AC power. Use only IVAC MedSystem III AC Adapter, Model 1550
Batteries:
Main – Rechargeable NiCad Battery Pack Memory
Back-up – Nonrechargeable Lithium
Battery charge:
A fully charged battery has a minimum of 6 hours running time with all channels running at 125 milliliter per hour and backlighting usage of 2 minutes per hour.The main battery retains 80% of its capacity after 500 charging Cycles, and retains 90% of its capacity after 3 months of Continuous AC charging.
AC Adapter & Cord Length:
Model 1550, 8.5 VDC @ 750 mA & 8.5 feet (2.59 meters)
AC Adapter Connector:
4 pin locking connector is standard on Model 2863. Detachable Connector is standard on Model 2860
Fuses:
3 amp fast-blow internal
Ground Continuity:
Maximum 0.1 ohm
Leakage Current:
Maximum 100 microamps