Medfusion 2010i
Product Specifications
Dimensions
4.5″ W x 3.0″ H x 7.5″ L
Weight
2.5 pounds
Accuracy
±3% excluding syringe variations
Infusion Modes
Body Weight modes:
– μg/kg/min μg/kg/min
– MG/kg/min MG/kg/HR
Mass Modes:
– μg/min μg/HR
– MG/min MG/HR
– mU/min mU/HR
– U/min U/HR
Continuous Mode:
– ml/min ml/HR
Volume Over Time Mode:
– Volume/time
Flow Rate
0.01 to 378.0 ml/HR dependent on syringe size selected
Bolus Delivery
In Body Weight, Mass and Continuous Modes
Power
AC Adapter – DC internal rechargeable batteries
Recharge Time
No longer than 16 hours with the pump turned off
Battery Capacity
at 25°C, a 16 hour charge will operate the pump for approximately 10 hours at 5.0 ml/hr with a 60 ml syringe.
Alarms/Alerts
Near Empty (5 to 60 minutes), Empty, Dose Volume Delivered, Bolus Complete, Bolus Delivery, Occlusion, System Malfunction, Low Battery, Depleted Battery, Syringe Pops Out, Invalid Size, Invalid Number, Check Clutch, Alarm Temporary Delay: 2 or 60 minutes, Alarm Audio Volume: Soft or Loud, Load Syringe Plunger
Status Alerts Visual, Non Audible
Stop/Program, Deliver, Battery In Use, Battery Charging, Battery Depleted, Priming, Standby Mode, Back Light
Total Delivery
From 0.0000 to 9999.9999 MG
From 0.00 to 9999.99 ml
From 0.0000 to 9999.9999 U
Syringe Options/Sizes
Becton-Dickinson (B-D®): 1,3,5,10,20,30,60 ml
Monoject®(Mono): 1,3,6,12,20,35,60 ml
Terumo® (Teru): 1,3,5,10,20,30,60 ml
Becton-Dickinson Glass (BG-G) 1,3,5,10 ml
Abboject®: 50 ml
Please note the following