Product Specifications
Dimensions
13.75″H x 12.2″W x 7.5″D
Weight
20 lbs. with batteries
Casing
High impact plastic
Delivery Rate Range
0.1-99.9 ml/hr in 0.1 ml increments100-999 ml/hr in 1 ml increments
Volume to be Infused Range
0.1-99.9 ml in 0.1 ml increments
100-9999 ml in 1 ml increments
KVO Rate
The lower of 1 ml/hr or the primary delivery rate
Occlusion Range
10 PSIGĀ (+5, -2 PSIG)
Power Requirements
100-130 VAC, 47/63 Hz, less than 60W
Power Cord
Hospital grade AC cord, 10ft long
Fuses
1.0 A, 250V, slow blowing
Batteries
Rechargeable and sealed, lead acid 8V batteries
Battery Life
A fully charged battery set will provide operating time of approximately 4 hours when all pumping units operate simultaneously at a rate of 125 ml/hr or a cumulative delivery of 1000 ml at any combination of rates and pumping units.
Battery Recharge
Recharge takes approximately 6 hours when the device is turned off.