CADD 5800 PCA

CADD 5800 PCA

Request Quote

Category:

Description

Pump Resolution:

50 microliters per interval

Dimensions:

1.1″ D x 3.5″ W x 6.4″ H including 50 ml reservoir

Weight:

Approximately 1 pound including battery and empty 50 ml reservoir

Power Source:

9-volt alkaline or lithium battery

Pump Timing Accuracy:

±0.1%

System Delivery Accuracy:

±6% nominal

Continuous Rate:

0-20 ml/hr or 0-99.5 mg/hr, depending on concentration

Patient Activated Dose Intervals:

5 min – 999 min in 1 min increments

Patient Activated Doses per Hour:

0-12 over the last 60 minute period

Accumulated Delivery:

0-999 mg or ml, rounded down according to concentration setting

Number of Patient Activated Doses Delivered:

0-999

Number of Patient Activated Doses Attempted:

0-999

Reservoir Volume Range:

1-999 ml rounded up to the nearest ml in 1 ml increments

Given Range:

0-999 mg or ml

Clinician Activated Bolus:

0-20 ml or 0-99.5 mg, depending on concentration

Patient Activated Dose:

0-6 ml or 0-99.0 mg, depending on concentration

Pump Alarms:

Low battery

Battery Depleted

Pump in Stop Mode

Controller, microprocessor, or motor fault

Improper Delivery

Power Up Fault

Low Residual volume

RES VOL = 000

High Delivery Pressure (28 ± 12 psi)