Description

Pump Resolution:

50 microliters per interval

Dimensions:

1.1″ D x 3.5″ W x 6.4″ H including 50 ml reservoir

Weight:

15 oz including battery and empty 50 ml reservoir

Power Source:

9 volt alkaline or lithium battery

Pump Timing Accuracy:

±0.1%

System Delivery Accuracy:

±6% nominal

Continuous Given Range:

0-999 ml in 1 ml increments rounded down to the nearest ml

Continuous RES VOL Alarm Range:

0-1500 ml in 1 ml increments rounded up to the nearest ml

Continuous ML/HR range:

0.00-75.0 ml/hr in 0.1 ml increments

Intermittent ML range:

0.00-150.0 ml in 0.1 ml increments

Intermittent Period Range:

10 min-12 hours, in 10 min increments

Intermittent Cycle Range:

20 min-12 hours and 10 min (20:10), in 10 min increments; and 24 hours for a 24 hour cycle

Intermittent KVO ml/hr Range:

0.2-10.0 ml/hr, in 0.1 ml increments

Intermittent Delayed Start Range:

10 min-12 hours, in 10 min increments

Intermittent Cycle Given Range:

0 min-12 hours and 10 min (12:10), in 1 min increments; and 0-24 hours in 1 hour increments when Cycle is programmed in 24 hour mode

Intermittent Given Range:

0-999 ml in 1 ml increments rounded down to the nearest ml

Intermittent RES VOL Alarm Range:

0-1500 ml in 1 ml increments rounded up to the nearest ml

Pump Alarms:

Low battery

Battery depleted

Pump in Stop Mode

Controller, microprocessor, or motor fault

Improper delivery

Power up fault

Low residual volume

High delivery pressure (22 ± 12 psi)